vol. 4, nr. 1 (2004)
|
Experiences with a TDR-moisture-meter in laboratory investigations |
|
|
Franz Feichtinger, Andreas Scheidl |
|
| (get PDF) |
|
|
Institute for Soil and Water Management, Pollnbergstrasse 1, A-3252 Petzenkirchen, Austria |
|
|
vol. 4 (2004), nr. 1,
pp. 37-41
|
|
|
abstract:
In a laboratory experiment, soil moisture was measured by means of a TDR-moisture-meter. Corresponding water content values were available from measurements of the gravimetric water content and the bulk density of the soils. Results for the period 1995-2000 confirmed the factory installed calibration functions and indicated that there was no long term drift of the TDR-system used.
|
|
keywords:
TDR, water content, soil, laboratory investigations
|
|
original in:
Polish
|
|
|
|
|